Mitutoyo step gauge 1um, 2,um, 5um, 10um
(1) Measurement result of difference and parallelism are described to inspection certificate.
(2) Differences are measured by interferometer measurement.
(3) Allowance of adjoined step difference against nominal value is less than 15%.
(4) Accuracy of measurement value of difference is less than 0.05um.
(5) Parallelism of each block against under surface is less than 0.05um (per 8mm).
(6) Gaps between blocks are less than 0.05mm.
(7) Chamfering of block is less than 0.05mm.